IoT-Based Remote Intravenous Infusion Monitoring and Alert System
This project, developed by students with equipment and component support from The Dariu Foundation, uses the Internet of Things (IoT) to monitor intravenous (IV) infusions from a distance and to alert nursing staff when something goes wrong. Optical and ultrasonic sensors, embedded control and an automatic tube-clamping mechanism work together at the bedside, while a central nursing station follows the readings wirelessly.
In busy hospitals, nurses cannot watch every infusion at the bedside all the time. A change in drip rate, an empty infusion container or air in the tubing creates a risk that needs timely attention. The team set out to build a system that measures each infusion, reports it remotely and responds to abnormal conditions.
HOW IT WORKS
The system works in sequence, from counting drops to intervening when a reading is abnormal:
- Infusion measurement and bedside display: A 20 mm slot-type optical sensor on the drip chamber counts drops. An ESP32-based controller estimates infusion volume (mL) and drip rate (drops per minute) and shows them at the bedside.
- Remote monitoring: The data reach a central nursing-station display over Wi-Fi. LoRa communication serves as an alternative connection when Wi-Fi is unavailable.
- Safety alerts and automatic intervention: An ultrasonic sensor is designed to detect an empty infusion line or air bubbles. The controller is then designed to sound an alarm, send a remote warning and close a motorised tube clamp.
- Portable power and hardware: An automatic UPS charging circuit and Li-ion batteries keep the system running when external power is disconnected. The hardware includes ESP32-S3 controller boards, LoRa modules and protective enclosures.
IMPACT
The system is meant to help nursing staff keep track of several infusions at once, identify abnormal conditions sooner and use their time more efficiently. The students’ work covered enclosure design and 3D printing, ESP32 firmware, a web-based monitoring system and sensor calibration. The safety functions require appropriate verification before clinical use.